semiatic
Our drawing for Mask fit result

Mask fit and mask seal results on a CPAP machine

Most machines can check the mask for you: the machine blows air while you wear the mask and it reports whether the seal is good. Some also give a verdict on last night’s seal in the morning summary. What each maker’s result means.

Therapy settings: This page explains what a CPAP machine’s screen or light is telling you, in the maker’s words. It is not medical advice: ask your equipment provider or clinician about your therapy settings, and do not change your pressure or any prescribed setting because of anything here.

ResMed

  1. ResMed AirSense 10 A morning summary screen

    Mask Seal (Sleep Report): “Good mask seal” or “Needs adjusting”

    After you stop therapy the Sleep Report shows one of two icons for Mask Seal, which the guide labels “Good mask seal” and “Needs adjusting, see Mask Fit”. It says how well the mask sealed during the session.

    What ResMed says to do

    1. If it needs adjusting, run Mask Fit: in My Options highlight Run Mask Fit and press the dial. The machine starts blowing air.
    2. Adjust the mask, the mask cushion and the headgear until you get a Good result. Press the dial or Start/Stop to stop Mask Fit.
    3. If you cannot get a good seal, talk to your care provider.

    Source: AirSense 10 user guide (AirSense 10 AutoSet, AutoSet for Her, Elite and CPAP, with HumidAir) (ResMed, 378947/1, 2021-02), checked 2026-09-29.

  2. ResMed AirSense 11 Something the machine does

    Mask Fit (MORE menu)

    A function that helps you assess and find possible air leaks around the mask. The Start Therapy/Standby light is white while it runs.

    What ResMed says to do

    1. For air leaking around the mask, ResMed says to make sure the mask is fitted correctly: see the mask user guide, run the Mask Fit function, or watch the Mask Fit video in the myAir app.

    Source: AirSense 11 user guide (Americas) (ResMed, 398133/2, 2021-10), checked 2026-09-29.

  3. ResMed AirMini Something the machine does

    Mask Fit (AirMini app)

    On the AirMini, Mask Fit runs from the AirMini phone app, not the machine.

    What ResMed says to do

    1. If air leaks from around the mask or the mask is noisy, make sure the mask is fitted correctly (see its user guide) or run Mask Fit in the app.
    2. Make sure the AirMini system is correctly assembled.

    Source: AirMini user guide (ResMed, 388135/1, 2020-03), checked 2026-09-29.

Everything else ResMed’s machines show: ResMed CPAP icons and messages.

Philips Respironics

  1. Philips Respironics DreamStation 2 A morning summary screen

    Last Session Mask fit: “Good fit” or “Adjust fit”

    A screen in the summary shown after therapy is turned off. “Good fit” means the leak found allows optimal performance of the device. “Adjust fit” means the leak may affect device performance; the machine still works and delivers therapy.

    What Philips Respironics says to do

    1. If “Adjust fit” shows, adjust the mask to reduce the leak before starting therapy again.

    Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.

  2. Philips Respironics DreamStation (first generation) An icon on the screen

    Check Mask Fit: green or red bar, then a green checkmark or a red “X”

    The optional Check Mask Fit (under My Setup, only if your home care provider has enabled it) measures the leak while the machine blows a test pressure and counts down 40 seconds. During the count a green bar means a good fit and a red bar means improvement is needed. Afterwards normal therapy starts and the screen shows a green checkmark (the leak allows optimal performance) or a red “X” (the leak may affect performance; the machine still works and delivers therapy).

    What Philips Respironics says to do

    1. If you want a better result, Philips says you can stop therapy, refit the mask and run Check Mask Fit again, following the instructions that came with your mask and headgear.

    Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.

  3. Philips Respironics DreamStation (first generation) A morning summary screen

    Mask Fit (My Info): a percentage

    “100% minus Large Leak”. Large Leak is the share of time the mask leak was so high that the machine could no longer identify breathing events with statistical accuracy. Shown for the last night and as averages over 7 and 30 days; only on the CPAP Pro and Auto CPAP models.

    The manual gives no step for this screen: the My Info screens only report, nothing on them can be edited.

    Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.

  4. Philips Respironics DreamStation Go An icon on the screen

    Check Mask Fit: a green checkmark or a red “X”

    The optional Check Mask Fit (under My Comfort, only if your healthcare provider has enabled it) checks the mask before therapy: the machine blows a test pressure while the screen counts down 40 seconds. A green checkmark means there is an appropriate amount of leak. A red “X” means the leak may affect performance; the machine still works and delivers therapy.

    What Philips Respironics says to do

    1. If you want a better result, Philips says you can stop therapy, refit the mask and run Check Mask Fit again, following the instructions that came with your mask and headgear.

    Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.

  5. Philips Respironics DreamStation Go A morning summary screen

    Mask Fit (My Info): a percentage

    “100% minus Large Leak”. Large Leak is the share of time the mask leak was so high that the machine could no longer identify breathing events with statistical accuracy. Shown for the last night and over the last 7 and 30 days.

    The manual gives no step for this screen; it only reports.

    Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.

Everything else Philips Respironics’s machines show: Philips Respironics CPAP icons and messages.

React Health

  1. React Health Luna G3 An icon on the screen

    Mask Fitting Test: “great” or “need to adjust”

    An option in the patient menu that tests whether the mask is worn correctly. The screen shows the “great” icon if it passes, otherwise the “need to adjust” icon.

    The manual gives no further step on this screen; its troubleshooting table points leaks from a wrong-size or badly positioned mask to your home care provider.

    Source: Luna G3 APAP / Luna G3 CPAP User Manual (React Health, V1.7, issue date February 29, 2024), checked 2026-09-29.

Everything else React Health’s machines show: React Health CPAP icons and messages.

Other things a CPAP machine shows

Sources

Each maker’s own user guide only, read on the date shown. Nothing here comes from third-party CPAP forums or “error code” sites.