What does this symbol on medical packaging mean?
Find it by its shape. 1. Pick what you see on the box, pouch or label. 2. Match the drawing. Each of the 85 symbols here has its own page with the standard’s wording, quoted from at least two publishers’ public documents.
Whether a device is right for you is a question for your pharmacist, nurse or doctor.
Tap a shape — the symbols that use it appear here.
A factory — 3 symbols
An hourglass — 1 symbol
Letters — 13 symbols
The word STERILE — 9 symbols
A crossed-out circle — 11 symbols
Limit lines — 3 symbols
Umbrella, sun or glass — 4 symbols
A triangle — 10 symbols
Warnings — 3 symbols
Arrows — 7 symbols
MR — 3 symbols
A book — 2 symbols
A rounded triangle — 7 symbols
Ovals — 4 symbols
Tubes, lines and valves — 5 symbols
A person or a heart — 12 symbols
A card, screen or calendar — 3 symbols
An alarm triangle or a bell — 6 symbols
Blue, red or yellow safety signs — 5 symbols
Marks on the device casing — 9 symbols
Reading a medical label: the four marks people ask about
- Hourglass + date is the use-by date. A date like 2027-08 runs to the end of that month. The outline factory with a date is the date of manufacture — not a deadline.
- LOT is the batch code a recall notice refers to; REF is the product code for reordering; SN is the serial number of one individual device.
- A crossed-out 2 means do not re-use: one use only. Several uses on the same person is a different mark, single patient – multiple use.
- STERILE with EO, R, A or a thermometer names the sterilisation method; a crossed-out torn box means the pack must be intact for that to hold.
Where these come from. ISO 15223-1:2021 “Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements” is a paid standard, so this page does not reproduce it. Each meaning is quoted as the cited public documents — manufacturers’ symbol glossaries, MedTech Europe’s guidance and US FDA and eCFR texts — reproduce it. The three MR marks come from ASTM F2503, which FDA’s MR labelling guidance follows, and “Rx only” from the US regulation 21 CFR 801.109. The applied-part marks (type B, BF, CF and their defibrillation-proof versions) are not ISO 15223-1 symbols: they are IEC 60417 graphical symbols that IEC 60601-1 requires on medical electrical equipment, so you find them on the device or its cable rather than on the box. The same holds for the marks printed on monitors, infusion pumps, oxygen concentrators and other medical electrical equipment: the alarm symbols (alarm, alarm off, alarm paused, audio off, audio paused, alarm reset) are IEC 60417 symbols used by IEC 60601-1-8, the alarm-systems standard, and IEC 60601-1 lists the rest in its Table D.1 (symbols such as equipotentiality) and Table D.2 (safety signs such as the blue follow instructions for use disc).
A heart with a line, a cuff icon or a coloured bar on the screen of a home blood pressure monitor? What blood pressure monitor symbols mean →
HI, LO, an apple or a thermometer on the screen of a blood glucose meter? What glucose meter symbols and messages mean →
Lo, Hi, Err or a red screen on a digital thermometer? What thermometer symbols and messages mean →
A number like ISO 7000-2607 or IEC 60417-5333 in a glossary or datasheet? Look up the symbol by its number →
SpO2, PR, PI, a moving wave or Finger Out on a fingertip pulse oximeter? What pulse oximeter readings and symbols mean →
What this page cannot do
- It cannot identify a product from its symbols or tell you who makes a device you found loose.
- It cannot decode what a lot number means for a given maker. Batch and serial codes are each maker’s own; only the maker can say what a code stands for or whether it is in a recall.
- It cannot say whether a device is safe or suitable for you, or whether an expired or damaged one is still usable. It reports what a symbol means under its standard, nothing more.
Left out, and why
- Control, negative control, positive control (CONTROL boxes) (ISO 15223-1:2021, 5.5.2 to 5.5.4): only NE Scientific’s glossary reproduced them until 2026-10-16, when QuidelOrtho’s symbol glossary was found to show all three boxes with the same wording, and Hologic’s glossary the minus and plus ones. Each now has its own page: Control (CONTROL box), Negative control (CONTROL −), Positive control (CONTROL +).
- Drops per millilitre and liquid filter with pore size (ISO 15223-1:2021, 5.6.4 and 5.6.5): only NE Scientific’s glossary among the documents we fetched reproduces them. Checked again on 2026-10-16 in the public glossaries of QuidelOrtho, Hologic, Vela Diagnostics, ICU Medical, Medela, Cook Medical, Integra, Stago, BÜHLMANN, Verathon, Biosense Webster, Ethicon, Globus Medical, Xtant Medical, LINK and Medtronic (MiniMed), and in MedTech Europe’s guidance: none prints either symbol. The symbol-glossary addresses we tried for B. Braun and Baxter returned “page not found” and BD’s refused the request, so those three makers remain unchecked. They stay out until a second publisher’s public document confirms the wording.
- Sampling site (ISO 15223-1:2021, 5.6.1): none of the eleven public glossaries we fetched reproduces its wording, so there is nothing to quote it from.
- Urgent alarm (triangle with two arcs) (IEC 60417-5308): only Hillrom’s glossary among the documents we fetched reproduces it. Checked again on 2026-10-16 in Mindray’s iPM patient-monitor operator’s manual and the glossaries of Biosense Webster, Ethicon, Integra, Verathon, Medela, ICU Medical, NE Scientific and Boston Scientific: none shows the two-arc triangle, so it stays out until a second publisher’s public document confirms it. The one-arc alarm triangle has its own page.
- Sound muting (crossed-out loudspeaker) (IEC 60417-5436): only Hillrom’s glossary among the documents we fetched reproduces it. Checked again on 2026-10-16 in Mindray’s iPM patient-monitor operator’s manual and the glossaries of Biosense Webster, Ethicon, Integra, Verathon, Medela, ICU Medical, NE Scientific and Boston Scientific: none shows the crossed-out loudspeaker under IEC 60417-5436.
- Nurse call (IEC 60417-5668): only NE Scientific’s glossary gives this IEC number (“Nurse”); Hillrom prints its own custom nurse-call symbol, and custom symbols differ from maker to maker.
- Stop (of action) (IEC 60417-5110): none of the equipment makers’ glossaries we fetched reproduces it.
- No sitting and No stepping on surface (ISO 7010-P018 and P019 (IEC 60601-1 Table D.2, safety signs 6 and 7)): No stepping on surface (P019) was published on 2026-10-16: Ethicon’s symbols glossary lists it beside Boston Scientific’s. No sitting (P018) stays out — only Boston Scientific’s glossary reproduces it; Ethicon’s lists P019 but not P018, and none of the other glossaries checked again on 2026-10-16 (Biosense Webster, Integra, Verathon, Medela, ICU Medical, NE Scientific, Hologic, QuidelOrtho, Globus Medical, Xtant Medical, LINK) shows it. See No stepping on surface (red prohibition sign).
Drawings semiatic already explains keep their one page: the Protective earth (IEC 60417-5019, IEC 60601-1, Table D.1, symbol 6), the Class II equipment (IEC 60417-5172, IEC 60601-1, Table D.1, symbol 9), the Stand-by (IEC 60417-5009, IEC 60601-1, Table D.1, symbol 29), the Dangerous voltage (lightning arrow) and the yellow “warning, electricity” sign (IEC 60417-5036, IEC 60601-1, Table D.1, symbol 24 and Table D.2, safety sign 3 (ISO 7010-W012)), the Alternating current (IEC 60417-5032, IEC 60601-1, Table D.1, symbol 1), the “ON”/“OFF” push-push (IEC 60417-5010, IEC 60601-1, Table D.1, symbol 14) — each has an “On medical equipment” section there.
Drawn twice on semiatic, with different meanings: Consult instructions for use (ISO 15223-1:2021, 5.4.3) and Book (read the operator’s manual); Temperature limit (ISO 15223-1:2021, 5.3.7 (with 5.3.5 and 5.3.6)) and Temperature limits (minimum and maximum °C). Each medical page says how the two differ.
Sources
Every document below is public and was fetched and checked on the date shown: manufacturers’ symbol glossaries, MedTech Europe’s guidance, and US FDA and eCFR texts. Each symbol page lists what its sources show.
- Boston Scientific: Symbols Glossary Checked 2026-09-25.
- NE Scientific: Master Symbol Glossary (CTF-100011 Rev 01) Checked 2026-09-25.
- BÜHLMANN Laboratories: Symbol glossary, version 03 (release date 2025-02-04) Checked 2026-09-25.
- MedTech Europe: Use of Symbols to Indicate Compliance with the MDR, version 3.0 (November 2021) Checked 2026-09-25.
- Stago: Glossary of symbols (520233-A_EN) Checked 2026-09-25.
- Cook Medical: Glossary of standard symbols Checked 2026-09-25.
- Integra LifeSciences: Symbols Glossary (1410819-2 Rev B) Checked 2026-09-25.
- SafeSource Direct: Medical Device Symbols Glossary Checked 2026-09-25.
- QuidelOrtho: QuidelOrtho Symbol Glossary Checked 2026-10-16.
- Hologic: Symbols Glossary (AW-16348 Rev. 003) Checked 2026-10-16.
- US Food and Drug Administration: Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment — guidance issued 10 October 2023 Checked 2026-09-25.
- eCFR (US Office of the Federal Register): 21 CFR 801.109 — Prescription devices Checked 2026-09-25.
- STERIS: STERIS Product Labeling, SCS-145, Exhibit A: Symbol Glossary Checked 2026-09-28.
- Hillrom: Symbols Glossary (80022945 Ver. J) Checked 2026-09-28.
- Ethicon: Symbols glossary (100584677 Rev.8) Checked 2026-09-28.
- Biosense Webster: BWI Symbols Glossary (FORM-3534 Rev C) Checked 2026-09-28.
- Mindray: iPM 12/iPM 10/iPM 8/iPM 7/iPM 6/iPM 5 Patient Monitor Operator’s Manual Checked 2026-09-28.
- ICU Medical: Symbols Glossary (IFU0000017 Rev 02) Checked 2026-09-28.
- Medela: Symbols Glossary (101046886 A) Checked 2026-09-28.
- Verathon: Symbol Glossary (0900-4757) Checked 2026-09-28.