Type CF applied part (heart in a square)

The standard’s wording, as the sources below reproduce it:

“To identify a type CF applied part complying with IEC 60601-1.”

IEC 60417-5335 · IEC 60601-1, Table D.1, symbol 21

What it means for you

An applied part is the part of the equipment that touches the patient — an electrode, a sensor, a probe. The letter is its protection class against electric shock under IEC 60601-1. Type CF (C for cardiac) has the strictest leakage-current limits and is the class for parts that may have direct contact with the heart, such as some cardiac catheters and the electrodes of equipment used during heart procedures.

Whether a device is right for you is a question for your pharmacist, nurse or doctor.

What it looks like
A filled heart inside a square.
Where you see it
On medical electrical equipment and its accessories — on the cable, the sensor, the connector socket or the rating plate — not on the sterile box. IEC 60601-1 Table D.1 lists it.
Other names
CF; heart in a box; cardiac floating applied part

What it does not tell you

It does not say the device is safe for you or suitable for a given procedure; it names the protection class only.

Often confused with

To identify a type BF applied part complying with IEC 60601-1.

More symbols in “A person or a heart”

Sources