Type CF applied part symbol, our drawing

Type CF applied part (heart in a square)

The standard’s wording, as the sources below reproduce it:

“To identify a type CF applied part complying with IEC 60601-1.”

IEC 60417-5335 · IEC 60601-1, Table D.1, symbol 21

B, BF or CF?

  • Type B — a standing figure, no frame (IEC 60417-5840): “Applied Part complying with specified requirements for protection against electric shock. Type B are those parts not suitable for direct cardiac application.”
  • Type BF — the figure inside a square (IEC 60417-5333): “F-Type applied part complying with a higher degree of protection against electric shock than type B applied parts. Type BF applied parts are those parts not suitable for direct cardiac application.”
  • Type CF — a heart inside a square (IEC 60417-5335): “F-Type applied part complying with the highest degree of protection against electric shock. Type CF applied parts are those parts suitable for direct cardiac application.” (this page)

Definitions quoted from EBME’s list of IEC 62353 / IEC 60601 terms (checked 2026-10-09, listed below); the drawings are described in our words.

What it means for you

An applied part is the part of the equipment that touches the patient — an electrode, a sensor, a probe. The letter is its protection class against electric shock under IEC 60601-1. Type CF (C for cardiac) has the strictest leakage-current limits and is the class for parts that may have direct contact with the heart, such as some cardiac catheters and the electrodes of equipment used during heart procedures.

Whether a device is right for you is a question for your pharmacist, nurse or doctor.

What it looks like
A filled heart inside a square.
Where you see it
On medical electrical equipment and its accessories — on the cable, the sensor, the connector socket or the rating plate — not on the sterile box. IEC 60601-1 Table D.1 lists it.
Other names
CF; heart in a box; cardiac floating applied part

What it does not tell you

It does not say the device is safe for you or suitable for a given procedure; it names the protection class only.

Often confused with

To identify a type BF applied part complying with IEC 60601-1.

More symbols in “A person or a heart”

Sources

IEC 60417 and IEC 60601-1 are paid standards, so the wording above is quoted as 4 publishers’ public documents reproduce it; our drawing is not the IEC artwork. Each was fetched and read on the date shown:

  • Boston Scientific: Symbols Glossary — IEC 60601-1 Table D.1, symbol 21 “Type CF applied part — To identify a type CF applied part complying with IEC 60601-1.” Checked 2026-09-28.
  • NE Scientific: Master Symbol Glossary (CTF-100011 Rev 01) — Shows the heart in a square for IEC 60417-5335, same description. Checked 2026-09-28.
  • Hillrom: Symbols Glossary (80022945 Ver. J) — Lists “Type CF applied part”, IEC 60417-5335, same description. Checked 2026-09-28.
  • Ethicon: Symbols glossary (100584677 Rev.8) — Lists “Type CF applied part”, IEC 60417 reference 5335. Checked 2026-09-28.
  • EBME: Medical Electronic Equipment (Electrical safety testing in accordance with IEC 62353), section ‘Commonly used terms and definitions in IEC 62353 / 60601’ — Defines each class: “Type B applied part — Applied Part complying with specified requirements for protection against electric shock. … Type B are those parts not suitable for direct cardiac application.” “Type BF applied part — F-Type applied part complying with a higher degree of protection against electric shock than type B applied parts. Type BF applied parts are those parts not suitable for direct cardiac application.” “Type CF applied part — F-Type applied part complying with the highest degree of protection against electric shock. Type CF applied parts are those parts suitable for direct cardiac application.” An F-type part is one “electrically isolated from ground and other parts of the medical equipment. i.e. floating.” Checked 2026-10-09.

Plain-language parts are in our own words; quotations are the sources’ wording.

Made with AI tools. · Privacy