Authorised representative (EC REP)
The standard’s wording, as the sources below reproduce it:
“Indicates the authorized representative in the European Community / European Union.”
ISO 15223-1:2021, 5.1.2
What the REP boxes on a medical label mean
- EC REP
The manufacturer’s authorised representative in the European Union: ISO 15223-1:2021 symbol 5.1.2, “Indicates the authorized representative in the European Community / European Union” as NE Scientific’s glossary reproduces it. The name and address beside it belong to a company inside the EU; under Regulation (EU) 2017/745, Article 11, a device whose maker is not established in a Member State “may only be placed on the Union market if the manufacturer designates a sole authorised representative”.
Source: NE Scientific; EUR-Lex (Publications Office of the EU), listed below.
- EU REP
Yes: same role, newer letters. The EU’s Medical Device Coordination Group (MDCG 2021-5 rev.1 Appendix, June 2026) says the amendment EN ISO 15223-1:2021/A1:2025 shows the representative “as EU REP instead of EC REP, removing any reference to the term European Community”, and that the change “is purely editorial in nature”. Boxes may carry either until 17 June 2031; from then MDCG says compliance “is granted only by the use of” the amended standard, though devices placed on the EU market before that date may still be sold.
Source: European Commission, Medical Device Coordination Group, listed below.
- UK REP / UKRP
The UK Responsible Person, the Great Britain counterpart. MHRA guidance says that “to place a device on the Great Britain market (England, Wales and Scotland), manufacturers based outside the UK must appoint a UK responsible person”, whose name and address must be on the label, outer packaging or instructions “in cases where the UKCA marking has been affixed”. A box with only the CE mark does not need it. The MHRA page names the role but shows no UK REP symbol of its own, so the exact letters are the maker’s choice.
Source: MHRA (GOV.UK), listed below.
- CH REP
The Swiss authorised representative. Swissmedic gives the symbol’s description as “Indicates the authorised representative in Switzerland”, says “the name and address of the authorised representative must appear adjacent to the symbol”, and that a P.O. box, e-mail address or telephone number alone is not enough. Instead of the symbol, a label may say “CH authorised representative”, “CH-REP” or “Authorised representative for Switzerland”.
Source: Swissmedic, listed below.
What it means for you
When the manufacturer is based outside the EU, EU law requires it to name a company inside the EU that acts for it with regulators. The name next to EC REP is that company. The 2021 edition widened the wording: Boston Scientific’s glossary reads “authorized representative in the identified jurisdiction”, so representatives for jurisdictions other than the EU use the same idea.
Whether a device is right for you is a question for your pharmacist, nurse or doctor.
Who has to name one and why
EU law defines the role in Regulation (EU) 2017/745 on medical devices, Article 2(32): an authorised representative is “any natural or legal person established within the Union who has received and accepted a written mandate from a manufacturer, located outside the Union, to act on the manufacturer's behalf in relation to specified tasks with regard to the latter's obligations under this Regulation”. Regulation (EU) 2017/746 on in vitro diagnostic devices (test kits, glucose strips) uses the same definition in Article 2(25). (Source: EUR-Lex (Publications Office of the EU).)
Article 11(1) of both regulations makes it compulsory for makers outside the EU: “Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative.” That is why a box from an EU-based maker has no EC REP at all. Annex I, 23.2(d) of 2017/745 then puts it on the label: “if the manufacturer has its registered place of business outside the Union, the name of the authorised representative and address of the registered place of business of the authorised representative”. (Source: EUR-Lex (Publications Office of the EU).)
What the named company has to do, from Article 11(3): keep the technical documentation and EU declaration of conformity “at the disposal of competent authorities”, “cooperate with the competent authorities on any preventive or corrective action”, and “immediately inform the manufacturer about complaints and reports from healthcare professionals, patients and users about suspected incidents related to a device for which they have been designated”. Under Article 11(5), if the maker has not met its own obligations, the representative “shall be legally liable for defective devices on the same basis as, and jointly and severally with, the manufacturer”. (Source: EUR-Lex (Publications Office of the EU).)
What you can do with it: the address beside the symbol is a real company in the EU (or Great Britain, or Switzerland) that regulators deal with for that maker, and Article 11(3)(g) obliges an EU representative to pass a reported incident on to the manufacturer immediately, so it is a usable contact when the maker itself is overseas. For Great Britain, MHRA says the UK Responsible Person’s “name and address are displayed within the manufacturer record” in its Public Access Registration Database (PARD), but “it is not currently possible to search PARD by Representative”: look the maker up by its own name. (Source: EUR-Lex (Publications Office of the EU); MHRA (GOV.UK).)
- What it looks like
- A rectangle split into two cells: EC on the left, REP on the right, with a company name and address next to it.
- Where you see it
- Boxes of devices made in the US, UK, China and elsewhere that are sold in the EU — for example test kits, contact lenses and diabetes supplies.
- Other names
- EC REP; EU REP; authorized representative; European representative
What it does not tell you
It does not say the product was made or tested in Europe, and the representative is not necessarily the seller. A product from an EU manufacturer has no EC REP at all.
Often confused with
Indicates the entity importing the medical device into the locale.
More symbols in “Letters”
Sources
ISO 15223-1 is a paid standard, so the wording above is quoted as 4 publishers’ public documents reproduce it. Each was fetched and read on the date shown:
- NE Scientific: Master Symbol Glossary (CTF-100011 Rev 01) — Shows the EC | REP box for reference no. 5.1.2, “Authorized Representative in the European Community/European Union — Indicates the authorized representative in the European Community / European Union”. Checked 2026-09-25.
- Boston Scientific: Symbols Glossary — ISO 15223-1 5.1.2 “Authorized representative — Indicates the authorized representative in the identified jurisdiction.” Checked 2026-09-25.
- Cook Medical: Glossary of standard symbols — ISO 15223-1 5.1.2 “Authorized Representative in the European Community/European Union — Indicates the Authorized Representative in the European Community/European Union.” Checked 2026-09-25.
- BÜHLMANN Laboratories: Symbol glossary, version 03 (release date 2025-02-04) — “This symbol shall be accompanied by the name and address of the authorized representative, adjacent to the symbol.” Checked 2026-09-25.
- EUR-Lex (Publications Office of the EU): Regulation (EU) 2017/745 on medical devices — Article 2(32), the definition of ‘authorised representative’; Article 11, “Authorised representative”, paragraphs 1, 3 and 5; Annex I, 23.2(d), the label particular. Checked 2026-10-09.
- EUR-Lex (Publications Office of the EU): Regulation (EU) 2017/746 on in vitro diagnostic medical devices — Article 2(25), the same definition of ‘authorised representative’; Article 11(1), the same duty to designate a sole authorised representative. Checked 2026-10-09.
- European Commission, Medical Device Coordination Group: MDCG 2021-5 rev.1 Appendix: Transition to the “EU REP” symbol in EN ISO 15223-1 (June 2026) — EU REP replaces EC REP in EN ISO 15223-1:2021/A1:2025; the change “is purely editorial in nature”; the transition runs until 17 June 2031. Checked 2026-10-09.
- MHRA (GOV.UK): Regulating medical devices in the UK (last updated 20 February 2026) — Section ‘UK responsible person’: who must appoint one, and when the name and address must be on the labelling. Checked 2026-10-09.
- MHRA (GOV.UK): Register medical devices to place on the market (last updated 20 July 2026) — Section ‘Public register of manufacturers’: PARD shows the UK Responsible Person within the manufacturer record and cannot be searched by representative. Checked 2026-10-09.
- Swissmedic: Swiss authorised representative (CH-REP) — Section ‘Use of the “CH-REP” symbol’: the symbol description, the name-and-address rule and the permitted wordings instead of the symbol. Checked 2026-10-09.
Plain-language parts are in our own words; quotations are the sources’ wording. Not covered here, because no source we opened says it: how to look an EU representative up in EUDAMED, and any official UK REP symbol drawing.