Philips Respironics CPAP machine icons, messages and lights
Philips Respironics makes three CPAP machines covered here, and their screens differ, so check which one you have before you match a message. The first-generation DreamStation (CPAP, CPAP Pro and Auto CPAP, recalled in 2021) has a control dial you turn and press, a Therapy button and a separate Ramp button on top; its alerts come with a blinking Power LED and sometimes a beep. The DreamStation 2 (Auto CPAP and CPAP) has a small touch screen and a Therapy button with a light ring. The DreamStation Go is the small travel machine with a touch screen you tap and swipe, an optional battery pack and a microSD card instead of a full-size SD card. Most of what each one tells you comes as a pop-up message.
Therapy settings: This page explains what a CPAP machine’s screen or light is telling you, in the maker’s words. It is not medical advice: ask your equipment provider or clinician about your therapy settings, and do not change your pressure or any prescribed setting because of anything here.
Philips Respironics DreamStation 2
23 entries from DreamStation 2 Auto CPAP / CPAP user manual (1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 A morning summary screen
Last Session Mask fit: “Good fit” or “Adjust fit”
A screen in the summary shown after therapy is turned off. “Good fit” means the leak found allows optimal performance of the device. “Adjust fit” means the leak may affect device performance; the machine still works and delivers therapy.
What Philips Respironics says to do
- If “Adjust fit” shows, adjust the mask to reduce the leak before starting therapy again.
More on this, and what other makers’ machines show: Mask fit result.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 An icon on the screen
Humidification symbol
Shown on the therapy screen when the adjustable humidifier setting is enabled. Tapping it shows the current setting: Off, Low, Medium or High.
No action is needed. The manual explains how to change the setting; ask your equipment provider which suits you.
More on this, and what other makers’ machines show: Humidifier and humidity icons.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Humidifier Error
Humidification control is disabled. There may be a problem with the humidifier; therapy will run without humidification.
What Philips Respironics says to do
- Turn off the device and disconnect it from power.
- Check by eye that the electrical contacts are clear, then reconnect the power cord.
- If the alert continues, contact your provider.
More on this, and what other makers’ machines show: Humidifier fault.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Humidifier Off
The attached battery does not support humidification, so the humidifier is off while you run from the battery pack.
What Philips Respironics says to do
- Disconnect the battery pack and use a compatible power supply, or use the device without humidification.
More on this, and what other makers’ machines show: Humidifier fault.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Tube Error
Heated tube error (only when a heated tube is fitted): the heated tube may be overheated or damaged.
What Philips Respironics says to do
- Turn off the device.
- Detach the heated tube, make sure it is not covered or obstructed, then reattach it.
- If the alert continues, contact your provider.
More on this, and what other makers’ machines show: Heated tube.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 An icon on the screen
RAMP PLUS (Ramp+) symbol
RAMP PLUS keeps a lower starting pressure for a set time (30 minutes by default) while you fall asleep. During that time the pressure stays at the starting level unless the machine detects an event and sees a need to increase it; afterwards your prescribed pressure resumes. While it runs, the pressure number on the therapy screen is the pressure being delivered at that moment.
No action is needed. The manual explains the setting; its levels are for you and your equipment provider to agree.
More on this, and what other makers’ machines show: Ramp.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 An icon on the screen
Cell modem signal strength
Shows the signal strength of the cellular modem on machines that have one. It is not shown in Airplane Mode.
The manual gives no step for a weak signal.
More on this, and what other makers’ machines show: Wireless or cellular signal.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 An icon on the screen
Airplane mode symbol
Airplane Mode is on. It replaces the cell and Bluetooth symbols while active, and it turns itself off automatically after three days.
No action is needed. The manual explains how to switch it in the device settings screens.
More on this, and what other makers’ machines show: Airplane mode.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 An icon on the screen
Bluetooth symbol, grey or white
Grey: Bluetooth is available. White: the machine is connected to a Bluetooth client (phone, tablet). Not shown in Airplane Mode. When a Bluetooth pulse oximeter is paired, the Bluetooth symbol and “SpO2” appear on the therapy screen.
No action is needed.
More on this, and what other makers’ machines show: Bluetooth.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
SD Error, Remove/Reinsert
The machine cannot read the SD card: the card may have a problem, it may have been removed while being written to, or it may have been inserted incorrectly.
What Philips Respironics says to do
- Press the Therapy button to clear the notification.
- Remove and reinsert the SD card.
- If the alert continues, contact your provider.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
SD Full, Contact provider
The SD card is full.
What Philips Respironics says to do
- Remove the SD card and replace it with a new one, or contact your provider for a new card.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
SD Removed, Reinsert
The SD card was removed and not put back before this therapy session started.
What Philips Respironics says to do
- Reinsert the SD card to record therapy data.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
SD Card Activity, Do not remove
A status message: the SD card is being read or written. No action is needed; it times out by itself.
What Philips Respironics says to do
- Do not remove the card while it shows.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Service Required, Contact support (light blinks continuously)
A device error has put the machine into a “Safe State”: the power stays on but the airflow is disabled.
What Philips Respironics says to do
- Disconnect the machine from power, then reattach the power cord.
- If the alert keeps coming back, contact your home care provider.
More on this, and what other makers’ machines show: System fault and error codes.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 A light on the machine
Therapy button ring: blue or white
Blue: therapy is on (the ring goes dark when therapy is on and the screen has gone dark). White: the machine is plugged in and therapy is off.
No action is needed. A ring that blinks continuously comes with a message: see Service Required, Check Power or Low Voltage.
More on this, and what other makers’ machines show: Light on the start button.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Check Power (light blinks continuously)
An incompatible (incorrect) power supply is attached.
What Philips Respironics says to do
- Switch to a Philips-provided power supply that can support therapy. Philips says an 80 W power supply is needed to support humidification.
More on this, and what other makers’ machines show: Power and voltage alerts.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Low Voltage (light blinks continuously)
Low voltage: an incompatible power supply is attached.
What Philips Respironics says to do
- Confirm a compatible Philips Respironics power supply is attached, and switch to one if needed.
- If you are using a battery, make sure it is adequately charged.
More on this, and what other makers’ machines show: Power and voltage alerts.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Blocked inlet
Blocked airway: something is blocking the air inlet on the machine.
What Philips Respironics says to do
- Check the air inlet is not obstructed.
- Check the air filter(s) are installed properly; replace them if needed.
More on this, and what other makers’ machines show: Tubing or air inlet blocked.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Low leak
Blocked airway: a blockage at the tube or the mask.
What Philips Respironics says to do
- Check the tube is not crushed or folded so that air flow is restricted.
- Check the mask is attached properly and nothing is obstructing it.
More on this, and what other makers’ machines show: Tubing or air inlet blocked.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Prescription Updated
A status message: your prescription has been updated. No action is needed.
No action is needed; press the Therapy button to clear it, or it times out.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Prescription Invalid
A prescription or setting update was unsuccessful.
What Philips Respironics says to do
- Contact your home care provider.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Words on the screen
Device Update, Hold to Confirm, Click to Cancel
A new software version is available for the machine.
What Philips Respironics says to do
- Press and hold the Therapy button to update the software, or click it to cancel.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation 2 Something the machine does
Airflow turns off by itself when you take the mask off
If your home care provider has enabled the Automatic Off feature, the machine turns the airflow off when you take off your mask and it recognises inactivity.
The DreamStation 2 manual describes the feature but lists no pop-up message for it. To carry on, put the mask back on and start therapy with the Therapy button.
More on this, and what other makers’ machines show: Automatic Off.
Source: DreamStation 2 Auto CPAP / CPAP user manual (Philips Respironics, 1143534 R02, 06/11/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation)
25 entries from DreamStation CPAP, CPAP Pro, Auto CPAP user manual (1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) An icon on the screen
Check Mask Fit: green or red bar, then a green checkmark or a red “X”
The optional Check Mask Fit (under My Setup, only if your home care provider has enabled it) measures the leak while the machine blows a test pressure and counts down 40 seconds. During the count a green bar means a good fit and a red bar means improvement is needed. Afterwards normal therapy starts and the screen shows a green checkmark (the leak allows optimal performance) or a red “X” (the leak may affect performance; the machine still works and delivers therapy).
What Philips Respironics says to do
- If you want a better result, Philips says you can stop therapy, refit the mask and run Check Mask Fit again, following the instructions that came with your mask and headgear.
More on this, and what other makers’ machines show: Mask fit result.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) A morning summary screen
Mask Fit (My Info): a percentage
“100% minus Large Leak”. Large Leak is the share of time the mask leak was so high that the machine could no longer identify breathing events with statistical accuracy. Shown for the last night and as averages over 7 and 30 days; only on the CPAP Pro and Auto CPAP models.
The manual gives no step for this screen: the My Info screens only report, nothing on them can be edited.
More on this, and what other makers’ machines show: Mask fit result.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Humidification Error. Contact support if the problem persists.
Humidifier error (only when a humidifier is fitted): a heater plate error, or the humidifier is not properly connected to the machine.
What Philips Respironics says to do
- Turn off the machine and disconnect it from power.
- Detach the humidifier, check by eye that the electrical contacts are clear, then reconnect the humidifier and the power cord.
- If the alert continues, contact your provider.
More on this, and what other makers’ machines show: Humidifier fault.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
The attached power supply does not support humidification.
The power supply plugged in cannot run the humidifier or the heated tube. Philips classes it as an Alert 2: pop-up, blinking Power LED and a beep, and it can happen during therapy.
What Philips Respironics says to do
- Switch to a Philips Respironics DreamStation power supply that can support humidification, or use the machine without the humidifier.
Philips’ troubleshooting table names the 80 W power supply as the one that runs the heated tube.
More on this, and what other makers’ machines show: Humidifier fault.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Heated Tube Error. Contact support if the problem persists.
Heated tube error (only when a heated tube is fitted): the tube may be overheated or damaged.
What Philips Respironics says to do
- Turn off the machine.
- Detach the heated tube from the humidifier, make sure the tube is not covered or obstructed, then reattach it.
- If the alert continues, contact your provider.
More on this, and what other makers’ machines show: Heated tube.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Something the machine does
Ramp button, and the green circle on the ramp screen
If your home care provider has enabled ramp, pressing the Ramp button on top of the machine after the airflow starts lowers the pressure while you fall asleep, then brings it up gradually to your prescription. The therapy screen shows the ramp pressure and a green circle fills as the pressure rises. Philips describes two modes, chosen by the provider: standard ramp rises at a steady rate; SmartRamp holds a lower pressure until the machine detects you need more.
What Philips Respironics says to do
- You can use the Ramp button as often as you wish during the night.
- If the button does nothing, Philips says ramp may not be prescribed for you, or the ramp starting pressure is already the same as the therapy pressure. Discuss it with your home care provider.
More on this, and what other makers’ machines show: Ramp.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Pair?: 123456 Yes/No
A phone or other Bluetooth device is asking to pair with the machine. The digits identify this machine.
What Philips Respironics says to do
- Turn the control dial to Yes to accept or No to decline, then press the dial to confirm.
More on this, and what other makers’ machines show: Bluetooth.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Data Activity: Do not remove SD card.
A status message: the SD card is being read or written.
No action is needed. Like every status message on this machine, it disappears by itself after 30 seconds.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
SD Card Removed.
The SD card was removed and not put back before this therapy session started. Philips lists it as a Notification (pop-up and blinking Power LED) or an Alert 2 (with a beep as well, and it can happen during therapy).
What Philips Respironics says to do
- Reinsert the SD card, or press the control dial to clear the alert.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
SD Card Error: Remove and Reinsert
The machine cannot read the SD card: the card may have a problem, it may have been ejected while being written to, or it was inserted incorrectly.
What Philips Respironics says to do
- Remove the SD card and reinsert it.
- If the alert keeps coming back, replace it with another card or contact your provider.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
SD Card Full.
The SD card is full.
What Philips Respironics says to do
- Remove the SD card and replace it with a new card, or contact your provider for a new SD card.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Service Required
A device error has put the machine into “Safe State”: the power stays on but the airflow is disabled. It comes with a blinking Power LED and a repeating beep.
What Philips Respironics says to do
- Press the control dial to silence the alert.
- Disconnect the machine from power, then reattach the power cord to restore power.
- If the alert continues, contact your home care provider.
More on this, and what other makers’ machines show: System fault and error codes.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) A light on the machine
Therapy On/Off button light flashing
You may have a message waiting to be read.
What Philips Respironics says to do
- Press or turn the control dial to show the message.
More on this, and what other makers’ machines show: Light on the start button.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) A light on the machine
Power LED blinking, with or without a beep
The machine is showing an alert. Philips sorts its pop-ups by what comes with them: Status (the pop-up only; it clears itself after 30 seconds), Notification (plus a blinking Power LED), Alert 1 (plus a beep; never during therapy), Alert 2 (plus a beep; can happen during therapy) and Safe State (plus a repeating beep).
What Philips Respironics says to do
- Read the pop-up and look it up on these pages. Every type except Status has to be acknowledged, usually by pressing the control dial, before it clears.
More on this, and what other makers’ machines show: Light on the start button.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Check Power
An incompatible power supply is attached, or the power cord is not fully pushed into the machine’s power inlet. The pop-up comes with a blinking Power LED.
What Philips Respironics says to do
- Make sure the power cord is fully inserted into the machine’s power inlet.
- Make sure a compatible Philips Respironics power supply is attached; switch to one if needed.
More on this, and what other makers’ machines show: Power and voltage alerts.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Low Voltage
Low voltage: an incompatible power supply is attached.
What Philips Respironics says to do
- Make sure a compatible Philips Respironics power supply is attached; switch to one if needed.
- If you are running from a battery, make sure it is adequately charged.
More on this, and what other makers’ machines show: Power and voltage alerts.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Something the machine does
Nothing happens when you plug it in; the button backlights stay dark
There is no power at the outlet, or the machine is unplugged.
What Philips Respironics says to do
- On mains power: check the outlet works and the machine is properly plugged in, the AC cord is connected to the power supply, and the power supply cord is firmly in the machine’s power inlet.
- On DC power: make sure the DC cord and battery adapter cable connections are secure and check the battery. If it still does not start, check the DC cord’s fuse as the cord’s instructions describe.
- If the problem continues, contact your home care provider and return both the machine and the power supply, so they can tell which one is at fault.
More on this, and what other makers’ machines show: Power and voltage alerts.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Inlet blocked. Check filter.
Blocked airway: something is blocking the air inlet on the machine.
What Philips Respironics says to do
- Check the machine’s air inlet is not obstructed.
- Check the air filter or filters are installed properly; replace them if needed.
More on this, and what other makers’ machines show: Tubing or air inlet blocked.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Low Leak: Check Mask and Tube
Blocked airway: a blockage at the tube or the mask.
What Philips Respironics says to do
- Check the tube is not crushed or folded so that airflow is restricted.
- Check the mask is attached properly and nothing is blocking it.
More on this, and what other makers’ machines show: Tubing or air inlet blocked.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Change Accepted
A status message confirming that a prescription change or a device software upgrade was accepted.
No action is needed. After a software update from an SD card, Philips says to remove the SD card to restart the machine on the new software.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Change Rejected
A prescription or settings change was rejected because it was missing or incorrect. Philips classes it as an Alert 1: pop-up, blinking Power LED and a beep; it does not happen during therapy.
What Philips Respironics says to do
- Contact your provider.
If it appears straight after a software update from an SD card, Philips says an SD card error was detected: remove the card and reinsert it, and if the alert continues, call Philips Respironics for a new SD card.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Would you like to upgrade software?
An SD card carrying a new software version has been inserted while therapy is off.
What Philips Respironics says to do
- Turn the control dial to Yes and press it to start the upgrade. The busy icon shows while it runs; do not remove power from the machine.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Filter reminder (every 30 days)
An automatic reminder, every 30 days, to check your filters and replace them as directed. It is only a reminder: the machine does not measure the filters or notice when one has been rinsed or replaced.
What Philips Respironics says to do
- Check the filters. The reusable blue pollen filter can be rinsed; the light-blue ultra-fine filter is disposable. An undamaged blue pollen filter must be in place for the machine to work properly.
More on this, and what other makers’ machines show: Replace filter, tubing or mask reminders.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Automatic Off
Therapy ended because of the automatic off function: the mask was removed. A status message, so it clears itself after 30 seconds.
What Philips Respironics says to do
- Put your mask back on, check it fits well, and turn the airflow back on to resume therapy.
More on this, and what other makers’ machines show: Automatic Off.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation (first generation) Words on the screen
Patient Message
A message from your provider. It is a Notification: the pop-up comes with a blinking Power LED.
What Philips Respironics says to do
- Read it, then press the control dial to acknowledge and clear it. Questions about what it says go to your provider.
More on this, and what other makers’ machines show: Message from your provider.
Source: DreamStation CPAP, CPAP Pro, Auto CPAP user manual (Philips Respironics, 1152070 R00, 08/27/2021), checked 2026-09-29.
Philips Respironics DreamStation Go
27 entries from DreamStation Go Auto CPAP / CPAP user manual (1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go An icon on the screen
Check Mask Fit: a green checkmark or a red “X”
The optional Check Mask Fit (under My Comfort, only if your healthcare provider has enabled it) checks the mask before therapy: the machine blows a test pressure while the screen counts down 40 seconds. A green checkmark means there is an appropriate amount of leak. A red “X” means the leak may affect performance; the machine still works and delivers therapy.
What Philips Respironics says to do
- If you want a better result, Philips says you can stop therapy, refit the mask and run Check Mask Fit again, following the instructions that came with your mask and headgear.
More on this, and what other makers’ machines show: Mask fit result.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go A morning summary screen
Mask Fit (My Info): a percentage
“100% minus Large Leak”. Large Leak is the share of time the mask leak was so high that the machine could no longer identify breathing events with statistical accuracy. Shown for the last night and over the last 7 and 30 days.
The manual gives no step for this screen; it only reports.
More on this, and what other makers’ machines show: Mask fit result.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Humidification Error
Humidifier error (only when the optional humidifier is fitted): a heater plate error, or the humidifier is not properly connected to the machine.
What Philips Respironics says to do
- Turn off the machine and disconnect it from power.
- Detach the humidifier, check by eye that the electrical contacts are clear, then reconnect the humidifier and the power cord.
- If the alert continues, contact your provider.
More on this, and what other makers’ machines show: Humidifier fault.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Humidifier and Battery Unsupported
Therapy was started with both the battery pack and the humidifier attached. The DreamStation Go cannot give therapy with both connected.
What Philips Respironics says to do
- Remove the battery to use therapy with humidification.
More on this, and what other makers’ machines show: Humidifier fault.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Humidifier Water Tank Inserted Incorrectly
The water tank is not making contact with the humidifier’s heater plate.
What Philips Respironics says to do
- Tap the screen to clear the error.
- Take the water tank out and put it back so that it is fully and properly connected.
More on this, and what other makers’ machines show: Humidifier fault.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Pair?: 123456 Yes/No
A phone or other Bluetooth device is asking to pair with the machine. The digits identify this machine.
What Philips Respironics says to do
- Swipe the screen to Yes to accept or No to decline, then tap to confirm.
- The pop-up times out after 30 seconds, and pairing is cancelled if you have not chosen Yes.
More on this, and what other makers’ machines show: Bluetooth.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Bluetooth LE Passkey
After you accept pairing, the machine shows a passkey to type into your phone.
What Philips Respironics says to do
- Enter the passkey on your phone to pair. The pop-up times out after 30 seconds and pairing is cancelled if the passkey is not used.
More on this, and what other makers’ machines show: Bluetooth.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Data Activity: Do not remove microSD card
A status message: the microSD card is being read or written.
No action is needed; the message clears when the card activity is finished.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
microSD card removed
The microSD card was removed and not put back before this therapy session started.
What Philips Respironics says to do
- Reinsert the microSD card, or click to clear the alert.
To take the card out, Philips says to turn off therapy, pull out the filter, and push the card in with the end of the filter so that it comes out.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
microSD card error: Remove and reinsert
The machine cannot read the microSD card: the card may have a problem, it may have been ejected while being written to, or it was inserted incorrectly.
What Philips Respironics says to do
- Remove the microSD card and reinsert it.
- If the message comes back, contact your healthcare provider for a replacement card.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
microSD card full
The microSD card is full.
What Philips Respironics says to do
- Remove the microSD card and replace it with a new card from your healthcare provider.
More on this, and what other makers’ machines show: SD card and data card.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Service Required
A device error has put the machine into “Safe State”: the power stays on but the airflow is disabled.
What Philips Respironics says to do
- Disconnect the machine from power, then reattach the power cord to restore power.
- If the alert continues, contact your healthcare provider.
More on this, and what other makers’ machines show: System fault and error codes.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Inlet blocked. Check filter.
Blocked airway: something is blocking the air inlet on the machine.
What Philips Respironics says to do
- Check the machine’s air inlet is not obstructed.
- Check the air filter or filters are installed properly; replace them if needed.
More on this, and what other makers’ machines show: Tubing or air inlet blocked.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Low leak: Check mask and tube
Blocked airway: a blockage at the tube or the mask.
What Philips Respironics says to do
- Check the tube is not crushed or folded so that airflow is restricted.
- Check the mask is attached properly and nothing is blocking it.
More on this, and what other makers’ machines show: Tubing or air inlet blocked.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Change Accepted
Confirms that a prescription change or a device upgrade was accepted.
What Philips Respironics says to do
- Tap the screen to acknowledge and clear the message.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Change Rejected
A prescription or settings change was rejected because it was missing or incorrect.
What Philips Respironics says to do
- Contact your healthcare provider.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Software Upgrade
The machine is asking whether to install a software update.
What Philips Respironics says to do
- Choose Yes or No. If you choose Yes, the upgrade is made: do not remove the machine from power. No clears the message.
More on this, and what other makers’ machines show: Prescription or settings update.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Filter reminder (every 30 days)
An automatic reminder, every 30 days, to check your filters and replace them as needed. It is only a reminder: the machine does not measure the filters or notice when one has been replaced.
What Philips Respironics says to do
- Check the filter. The reusable filter is washable; the disposable fine filter must not be rinsed. One of the two must be in place at all times to run the machine.
More on this, and what other makers’ machines show: Replace filter, tubing or mask reminders.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
Automatic Off
Therapy ended because of the automatic off function: the mask was removed.
What Philips Respironics says to do
- Put your mask back on, check it fits well, and turn the airflow back on to resume therapy.
More on this, and what other makers’ machines show: Automatic Off.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go Words on the screen
User Message
A message from your healthcare provider.
What Philips Respironics says to do
- Read it, then tap the screen to acknowledge and clear it. Questions about what it says go to your provider.
More on this, and what other makers’ machines show: Message from your provider.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go An icon on the screen
Charge bars in the lower right corner, during therapy
The optional battery pack’s charge while it is attached and therapy is running. Four white bars means fully charged; bars disappear as the charge drops. Philips says a fully charged pack typically lasts 8 hours, depending on your settings, mask leak and the surroundings.
No action: it is information. For how long the pack will last for you, Philips says to speak with your healthcare provider.
More on this, and what other makers’ machines show: Battery pack icons and lights.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go An icon on the screen
Battery percentage, therapy off
When the pack is attached but therapy is not running, the screen shows the charge as a percentage among the main menu screens.
No action: it is information.
More on this, and what other makers’ machines show: Battery pack icons and lights.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go An icon on the screen
Battery with a question mark
Battery Pack State of Charge Alert: the machine cannot work out how much charge is left.
The manual gives no step for this alert.
More on this, and what other makers’ machines show: Battery pack icons and lights.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go An icon on the screen
Battery with an X
Battery Pack Fault Alert: a fault has been detected in the battery pack.
The manual gives no step on the alert itself; its troubleshooting advice for a pack that may be damaged is under the rapidly flashing light below.
More on this, and what other makers’ machines show: Battery pack icons and lights.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go A light on the machine
Green light on the battery pack: steady, blinking or off
The pack’s single green light shows its charge while it charges on its own, not attached to the machine. Steady: fully charged. Blinking: charging. Off: the pack is connected to the therapy machine.
No action: it is a status light.
More on this, and what other makers’ machines show: Battery pack icons and lights.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go A light on the machine
Battery pack light flashing rapidly
Philips says the battery pack may have been damaged.
What Philips Respironics says to do
- If the pack has been very hot or very cold, let it come back to room temperature.
- Unplug the power cord from the pack, then plug it back in.
- If the light keeps flashing rapidly, replace the battery pack.
More on this, and what other makers’ machines show: Battery pack icons and lights.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
Philips Respironics DreamStation Go A light on the machine
Battery pack light does not come on while charging
Philips says the battery pack may have been damaged.
What Philips Respironics says to do
- If the pack was completely flat, wait a few minutes for the light to come on.
- If the pack has been very hot or very cold, let it come back to room temperature.
- If the light still does not come on, replace the battery pack.
More on this, and what other makers’ machines show: Battery pack icons and lights.
Source: DreamStation Go Auto CPAP / CPAP user manual (Philips Respironics, 1152853 R01, 06/14/2022), checked 2026-09-29.
What is not here
- Bilevel machines (Philips DreamStation BiPAP, ResMed AirCurve, Luna G3 BPAP) — Bilevel machines have their own screens, alarms and manuals; these pages cover CPAP and auto-CPAP machines only. www.documents.philips.com Checked 2026-09-29.
- Philips DreamStation (first generation) and DreamStation Go: status screens and oximetry — Time (a prompt to set the clock), Busy, Loading Language and Rebooting, Sleep Progress, the first-generation machine’s Oximetry icons (for the optional finger sensor) and its EZ-Start message are in the two manuals but not listed here: they need no action, or the message states a pressure, which these pages never do. Everything else in both manuals’ pop-up tables is. www.philips.com Checked 2026-09-29.
Sources
Philips Respironics’s own user guides only, read on the date shown.
- Philips Respironics: DreamStation 2 Auto CPAP / CPAP user manual — 1143534 R02, 06/11/2021. www.documents.philips.com Checked 2026-09-29.
- Philips Respironics: DreamStation CPAP, CPAP Pro, Auto CPAP user manual — 1152070 R00, 08/27/2021. www.philips.com Checked 2026-09-29.
- Philips Respironics: DreamStation Go Auto CPAP / CPAP user manual — 1152853 R01, 06/14/2022. www.philips.com Checked 2026-09-29.