Hi on a thermometer (above the measuring range)
The thermometer measured something hotter than the highest temperature it is built to report, so it shows Hi instead of a number. The manuals read treat it as a reading to take again after checking how it was taken, not as a temperature.
Check your thermometer’s manual — makers differ.
What it is for
Measure again, following your manual’s steps; Braun (ThermoScan) and Microlife (NC 200) give the same fix for Hi as for Lo: clean the probe or sensor and take the reading again. Omron’s ET720 manual says: “Use the unit within its measurable range. If a malfunction still exists, please contact OMRON customer service.” About a high temperature itself, Omron’s manual adds: “A high or prolonged fever requires medical attention, especially for small children. Please contact your doctor.”
These pages report what each maker’s manual says and quote it wherever health is involved. They do not diagnose anything, they are not medical advice, and they do not tell you whether a temperature is a fever or what to do about it; only your own thermometer’s manual and your doctor or nurse can do that for the person you are looking after.
- What it looks like
- The letters Hi, HI or just H where the number would be. Some makers show a word or an error screen instead.
- Names makers use
- Hi (Omron ET720); HI (Braun ThermoScan); H (Microlife IR 310, NC 200 and MT 410); Error (Withings Thermo); An error screen (Beurer FT 90 and FT 95, Braun No touch BNT 400)
Where makes differ
The ceiling depends on the maker and the mode. Omron’s ET720 manual shows Hi above 108.0°F (42.2°C) on the forehead and above 176°F (80°C) in surface mode. Braun’s ThermoScan manual shows HI above 42.2 °C (108 °F), and Beurer’s FT 95 manual gives its error above 42.2 °C (108 °F) on the forehead and above 80 °C in object mode. Microlife uses a plain H: its manuals give it above 43 °C on the IR 310 and in body mode on the NC 200, above 42.9 °C on the MT 410 stick thermometer, and above 99.9 °C in the NC 200’s object mode. Withings’ Thermo shows the word Error above 43.2°C (109.76°F). Braun’s No touch BNT 400 uses one message for both ends, shown above 43 °C or below 34 °C.
What it does not mean
Hi is not the fever alarm: the thermometers read here show a fever as a number with a colour or a symbol, and Hi only when the reading is off the scale. It is not AH either, which Microlife’s manuals define as the room being too hot.
Other symbols with the same shapes
Sources
No published standard we could check defines the symbols on a digital thermometer’s display, so this meaning is taken from 5 makers’ own manuals and support pages, each fetched and read on the date shown:
- Omron: Omron No-Touch Forehead Thermometer ET720 (MC-720-Z) — instruction manual (US, 5681249-5D, PDF) — Taking a Reading: “If the result is higher than 108.0°F (42.2°C), Hi appears on the display”; error table: surface measurement mode higher than 176°F (+80°C); Important Safety Information on high or prolonged fever. Checked 2026-09-29.
- Braun: Braun ThermoScan 7 (IRT 6520) and ThermoScan 3 (IRT 6030) infrared ear thermometers — instruction manual (Asia-Pacific, PDF) — Errors and troubleshooting: “Temperature taken is not within typical human temperature range (34 – 42.2 °C or 93.2 – 108 °F). HI = too high”; clean the probe tip and lens, fit a new lens filter and take a new temperature. Checked 2026-09-29.
- Braun: Braun No touch + forehead thermometer BNT 400 — instruction manual (Asia-Pacific, PDF) — Errors and troubleshooting: one message “when the measured temperature is lower than 34 °C (93.2 °F) or higher than 43 °C (109.4 °F)” or the room is outside 15 °C – 40 °C; “Re-measure the temperature”. Checked 2026-09-29.
- Microlife: Microlife NC 200 non-contact thermometer — instruction manual (IB NC 200-e N-V10, 2026-08-17, PDF) — Error messages: «H» = too high, higher than 43 °C / 109.4 °F in body mode or 99.9 °C / 211.8 °F in object mode; clean the sensor and measure again. Checked 2026-09-29.
- Microlife: Microlife MT 410 digital stick thermometer — instruction manual (IB MT 410, revision 2025-07-03, PDF) — Technical specifications: measurement range 32.0°C to 42.9°C; Temp. > 42.9°C: display «H». Checked 2026-09-29.
- Microlife: Microlife Ear Thermometer IR 310 — instruction manual (IB IR 310 N-V10, revision 2024-09-13, PDF) — Error Messages: Measured temperature too high, «H» when the measured temperature is higher than 43 °C / 109.4 °F. Checked 2026-09-29.
- Beurer: Beurer FT 95 non-contact thermometer — instructions for use (2021-06-22, PDF) — Error messages: forehead temperature mode higher than 42.2 °C (108 °F); object temperature mode higher than 80 °C (176 °F); contact the retailer or Customer Services if it repeats. Checked 2026-09-29.
- Withings: Withings Thermo — regulatory information and troubleshooting (PDF) — Troubleshooting: “Error” — the temperature is outside ranges, higher than 43.2°C (109.76°F); make a new measurement, referring to the user manual. Checked 2026-09-29.
Written in our own words; quotations are the makers’ wording.